Evaluation of transpulmonary ultrasound dilution cardiac output in piglets: accuracy, precision and trending ability with room temperature injectate

Published:November 25, 2022DOI:https://doi.org/10.1016/j.vaa.2022.11.008

      Abstract

      Objective

      Transpulmonary ultrasound dilution (TPUD) is a minimally invasive technique to measure cardiac output (CO) using a 1 mL kg–1 isotonic 37 °C saline injectate indicator. The objective was to evaluate the performance of TPUD using a room temperature saline injectate.

      Study design

      Prospective experimental trial.

      Animals

      A total of seven anesthetized male Yorkshire piglets.

      Methods

      Piglets aged 1 month and weighing 7.7–9.0 kg were anesthetized with detomidine–ketamine–hydromorphone–isoflurane and a pulmonary artery flow probe (PAFP) placed via a median sternotomy. The thoracic cavity remained open during measurement of CO by PAFP and TPUD. The TPUD indicators of 1 mL kg–1 0.9% saline at 37 °C and 20 °C were compared during infusions of phenylephrine and dobutamine, blood withdrawal and replacement. Bias, limits of agreement (LoAs) and percentage error (PE) between each iteration of PAFP and TPUD were measured with Bland–Altman plots. Trending ability via concordance, angular bias and radial LoA were compared.

      Results

      Bland–Altman plots showed negligible bias with varying LoAs. PEs of 22% and 38% were found for 37 °C and 20 °C saline injectates, respectively. In the four-quadrant plots, the concordance rate was 94% and 100% for measurements obtained with 37 °C and 20 °C saline injectates, respectively. Angular bias for both were < ±5 °, with radial LoA < ±7 °.

      Conclusions

      TPUD was accurate when using 1 mL kg–1 of isotonic saline at 37 °C in a range of CO within 0.2–0.8 L minute–1, and it reliably tracked positive and negative changes in CO. Room temperature (20 °C) indicator was less accurate but equally able to track direction of changes in CO.

      Clinical relevance

      The use of room temperature injectates allows an easy, readily available clinical application of TPUD CO monitoring while preserving the trending ability of the monitor.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Veterinary Anaesthesia and Analgesia
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Bland J.M.
        • Altman D.G.
        Statistical methods for assessing agreement between two methods of clinical measurement.
        Lancet. 1986; 1: 307-310
        • Boehne M.
        • Baustert M.
        • Paetzel V.
        • et al.
        Determination of cardiac output by ultrasound dilution technique in infants and children: a validation study against direct Fick principle.
        Br J Anaesth. 2014; 112: 469-476
        • Critchley L.A.H.
        • Critchley J.A.J.H.
        A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques.
        J Clin Monit Comput. 1999; 15: 85-91
        • Critchley L.A.
        • Lee A.
        • Ho A.M.
        A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output.
        Anesth Analg. 2010; 111: 1180-1192
        • Crittendon III, I.
        • Dreyer W.J.
        • Decker J.A.
        • Kim J.J.
        Ultrasound dilution: an accurate means of determining cardiac output in children.
        Pediatr Crit Care Med. 2012; 13: 42-46
        • Darling E.
        • Thuramalla N.
        • Searles B.
        Validation of cardiac output measurement by ultrasound dilution technique with pulmonary artery thermodilution in a pediatric animal model.
        Pediatr Cardiol. 2011; 32: 585-589
        • de Boode W.P.
        • van Heijst A.F.J.
        • Hopman J.C.W.
        • et al.
        Cardiac output measurement using an ultrasound dilution method: a validation study in ventilated piglets.
        Pediatr Crit Care Med. 2010; 11: 103-108
        • Evonuk E.
        • Imig C.J.
        • Greenfield W.
        • Eckstein J.W.
        Cardiac output measured by thermal dilution of room temperature injectate.
        J Appl Physiol. 1961; 16: 271-275
        • Krivitski N.M.
        Novel method to measure access flow during hemodialysis by ultrasound velocity dilution technique.
        ASAIO J. 1995; 41: M741-M745
        • Krivitski N.M.
        • Kislukhin V.V.
        • Thuramalla N.V.
        Theory and in vitro validation of a new extracorporeal arteriovenous loop approach for hemodynamic assessment in pediatric and neonatal intensive care unit patients.
        Pediatr Crit Care Med. 2008; 9: 423-428
        • Lemson J.
        • de Boode W.P.
        • Hopman J.C.W.
        • et al.
        Validation of transpulmonary thermodilution cardiac output measurement in a pediatric animal model.
        Pediatr Crit Care Med. 2008; 9: 313-319
        • Lindberg L.
        • Johansson S.
        • Perez-de-Sa V.
        Validation of an ultrasound dilution technology for cardiac output measurement and shunt detection in infants and children.
        Pediatr Crit Care Med. 2014; 15: 139-147
        • Omenås I.N.
        • Tronstad C.
        • Rosseland L.A.
        Accuracy and trending abilities of finger plethysmographic blood pressure and cardiac output compared to invasive measurements during caesarean delivery in healthy women: an observational study.
        BMC Anesthesiol. 2020; 20: 157
        • Perrino Jr, A.C.
        • Harris S.N.
        • Luther M.A.
        Intraoperative determination of cardiac output using multiplane transesophageal echocardiography: a comparison to thermodilution.
        Anesthesiology. 1998; 89: 350-357
        • Saugel B.
        Tracking changes in cardiac output: statistical considerations on the 4-quadrant plot and the polar plot methodology: erratum.
        Anesth Analg. 2015; 121: 1404
        • Saugel B.
        • Grothe O.
        • Wagner J.Y.
        Tracking changes in cardiac output: statistical considerations on the 4-quadrant plot and the polar plot methodology.
        Anesth Analg. 2015; 121: 514-524
        • Shih A.
        • Giguère S.
        • Vigani A.
        • et al.
        Determination of cardiac output by ultrasound velocity dilution in normovolemia and hypovolemia in dogs.
        Vet Anaesth Analg. 2011; 38: 279-285
        • Thiele R.H.
        • Bartels K.
        • Gan T.J.
        Cardiac output monitoring: a contemporary assessment and review.
        Crit Care Med. 2015; 43: 177-185
        • Tsutsui M.
        • Matsuoka N.
        • Ikeda T.
        • et al.
        Comparison of a new cardiac output ultrasound dilution method with thermodilution technique in adult patients under general anesthesia.
        J Cardiothorac Vasc Anesth. 2009; 23: 835-840
        • Van Wyk L.
        • Gupta S.
        • Lawrenson J.
        • de Boode W.P.
        Accuracy and trending ability of electrical biosensing technology for non-invasive cardiac output monitoring in neonates: a systematic qualitative review.
        Front Pediatr. 2022; 10851850
        • Vrancken S.L.
        • de Boode W.P.
        • Hopman J.C.
        • et al.
        Influence of lung injury on cardiac output measurement using transpulmonary ultrasound dilution: a validation study in neonatal lambs.
        Br J Anaesth. 2012; 109: 870-878